Design freeze, verification and the quality system
Before a device can be tested for submission, its design is frozen and proven to meet its own requirements, under a quality system an auditor could inspect.
Device development runs under design controls: written user needs, design inputs (requirements), design outputs (the actual design) and a design history file that records how one became the other. The FDA's quality system rule now incorporates the international standard ISO 13485, which is why investors ask whether you have a QMS "to 13485".
Design freeze is the moment the design is locked. From then on a change goes through formal change control, because every test that follows is only valid for the design it tested.
Verification asks whether you built the device right: bench testing against each requirement, biocompatibility, electrical safety and electromagnetic compatibility, sterilisation and shelf life, and software verification. Validation asks whether you built the right device: that it meets user needs, including human factors testing with real users. Together they are V&V.
A device seed round usually buys the prototype and the pathway; the Series A usually buys design freeze, V&V and submission readiness. Slipping design freeze is the most common reason a device raise runs short, because every later date moves with it.