Life sciences

The device roads: 510(k), De Novo and PMA

Short answer

Most devices reach the US market by showing they match one already sold (510(k)), by being classified as a new low-to-moderate risk type (De Novo), or by proving safety and effectiveness outright (PMA).

VCTerminal fundraising guide · Updated October 2026 · General information, not legal or financial advice. Your circumstances can change the answer.

Devices are classed by risk. Class I (a tongue depressor) is mostly exempt from review. Class II (an infusion pump, most diagnostic imaging, much software) usually goes through a 510(k). Class III (an implanted defibrillator, a heart valve) needs a PMA.

A 510(k), a premarket notification, argues substantial equivalence to a legally marketed device called the predicate. It rarely needs a clinical trial. The FDA says a 510(k) device is cleared, not approved. If there is no predicate but the risk is low to moderate, the De Novo pathway classifies the new device type; the FDA says the request is granted, and the device then becomes a predicate for others.

A Premarket Approval (PMA) is for Class III devices and rests on the company's own clinical evidence, usually a pivotal study run under an IDE. It is the only road where the FDA says approved, and it is the longest and most expensive.

Those review goals are counted in FDA days, and the clock stops every time the FDA is waiting on your answers, so the calendar runs longer. Measured from receipt to decision in the FDA's own data, a successful 510(k) took a median of about four to five months in 2025, and a De Novo or an original PMA about ten to eleven months in 2023-2025. Plan on the calendar, not the goal.

Picking the road is the first milestone investors ask about, because it sets the cost and time of everything after. Confirm it early in a Pre-Submission (Q-Sub), where the FDA gives written feedback on your pathway, testing and any study. The Breakthrough Devices Program adds priority interaction for devices that may better treat a serious condition. In Europe the device instead needs a CE mark from a notified body under the EU Medical Device Regulation (MDR).